Response of quartz crystal microbalance loaded with single-drop liquid in gas phase

1Citations
Citations of this article
8Readers
Mendeley users who have this article in their library.

Abstract

The frequency response of quartz crystal microbalance loaded by single-drop liquid is studied in this paper. Previous studies have shown that the relationship between resonant frequency and properties of liquid by completely immersing one side of the crystal in liquid. In this work, only localized portion of crystal was wetted by liquid droplet. Repeated experiment shows the relationship between liquid property include viscosity and density to resonant frequency. Furthermore, Theoretical formula describing the frequency change of the quartz crystal microbalance with liquid property is proposed. The predicted results showed distinct coincide with experimental results.

Cite

CITATION STYLE

APA

Yang, L., & Huang, X. (2014). Response of quartz crystal microbalance loaded with single-drop liquid in gas phase. Open Electrical and Electronic Engineering Journal, 8(1), 197–201. https://doi.org/10.2174/1874129001408010197

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free