Abstract
A simple, rapid, and precise RP-HPLC method for simultaneous analysis of atorvastatin calcium, metformin hydrochloride, and glimepiride in bulk and its pharmaceutical formulations has been developed and validated. These drugs were separated by using Grace Smart Altima C-8 column (250 × 4.6 mm, 5- μ m) with a mobile phase consisting of acetonitrile : phosphate buffer (60 : 40 (v/v), pH 3.0) at a flow rate of 1 mL/min, injection volume 25 µ L, and detection at 235 nm. Metformin, atorvastatin, and glimepiride were eluted with retention times of 2.57 min, 7.06 min, and 9.39 min, respectively. The method was validated for accuracy, precision, linearity, specificity, and sensitivity in accordance with ICH (Q2B) guidelines. The results of all the validation parameters were found to be within the acceptable limits. The calibration plots were linear over the concentration ranges from 10 to 150 µ g/mL, 20 to 200 µ g/mL, and 10 to 150 µ g/mL for atorvastatin, metformin, and glimepiride, respectively. The accuracy and precision were found to be between 98.2%–105% and ≤2% for three drugs. Developed method was successfully applied for the determination of the drugs in tablet dosage form and recovery was found to be >98% for three drugs. The degradation products produced as a result of stress studies did not interfere with drug peaks.
Cite
CITATION STYLE
Ramesh, D., & Habibuddin, M. (2014). Stability Indicating RP-HPLC Method for the Simultaneous Determination of Atorvastatin Calcium, Metformin Hydrochloride, and Glimepiride in Bulk and Combined Tablet Dosage Form. International Scholarly Research Notices, 2014, 1–8. https://doi.org/10.1155/2014/754695