Plasma L-Ergothioneine Measurement by High-Performance Liquid Chromatography and Capillary Electrophoresis after a Pre-Column Derivatization with 5-Iodoacetamidofluorescein (5-IAF) and Fluorescence Detection

31Citations
Citations of this article
32Readers
Mendeley users who have this article in their library.

Abstract

Two sensitive and reproducible capillary electrophoresis and high-performance liquid chromatography-fluorescence procedures were established for quantitative determination of L-egothioneine in plasma. After derivatization of L-ergothioneine with 5-iodoacetamidofluorescein, the separation was carried out by HPLC on an ODS-2 C-18 sperisorb column by using a linear gradient elution and by HPCE on an uncoated fused silica capillary, 50 μm id, and 60 cm length. The methods were validated and found to be linear in the range of 0.3 to 10 μmol/l. The limit of quantification was 0.27 μmol/l for HPCE and 0.15 μmol/l for HPLC. The variations for intra- and inter-assay precision were around 6 RSD%, and the mean recovery accuracy close to 100% (96.11%). © 2013 Sotgia et al.

Cite

CITATION STYLE

APA

Sotgia, S., Pisanu, E., Pintus, G., Erre, G. L., Pinna, G. A., Deiana, L., … Zinellu, A. (2013). Plasma L-Ergothioneine Measurement by High-Performance Liquid Chromatography and Capillary Electrophoresis after a Pre-Column Derivatization with 5-Iodoacetamidofluorescein (5-IAF) and Fluorescence Detection. PLoS ONE, 8(7). https://doi.org/10.1371/journal.pone.0070374