Impact of haematocrit on the accuracy of spot blood glucose measurements in dogs using two veterinary hand-held glucometers

1Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.

Abstract

Anaemia is a common finding in critically ill dogs, and understanding its potential impact on the accuracy of point-of-care glucometers is essential. In human medicine, it is well reported that haematocrit can affect blood glucose measurements and influence clinical outcomes when measured using certain hand-held glucometers, whereas evidence in veterinary medicine is still scarce. This prospective case–control study included 72 client-owned dogs, 39 of which were anaemic. Haematocrit values were used to classify dogs into anaemic and non-anaemic categories. Whole blood glucose concentrations were measured using two new models of veterinary hand-held glucometers (G-PET PLUS® and AlphaTrak3®) and compared with the serum blood glucose measurements obtained with an automated wet biochemistry analyser. Agreement between the point-of-care blood glucose analysers and the reference method, as well as the effects of haematocrit on measurement accuracy, was assessed. Both hand-held glucometers had a significant positive correlation with the wet biochemistry analyser (AlphaTrak3®: r = 0.664; 95% CI 0.502, 0.78; G-PET PLUS®: r = 0.769; 95% CI 0.643, 0.855). However, when comparing the overall blood glucose results, there was a significant negative percentage bias (p < 0.001) for both devices. Haematocrit reductions did not affect the degree of bias for AlphaTrak3®, and for the G-PET PLUS®, bias was also present in control patients and in patients with a mild haematocrit reduction. Despite these statistical differences, clinical impact assessments of the results via specialist clinician questionnaires and Parkes error grid analysis showed no change in clinical decision-making or patient outcome, supporting the cautious use of these devices in anaemic but normoglycaemic dogs when testing whole blood EDTA samples.

Cite

CITATION STYLE

APA

Garcia, M. R., Mann, N., Perelló, M. H., Silic, V., Sparks, T., & Titmarsh, H. (2025). Impact of haematocrit on the accuracy of spot blood glucose measurements in dogs using two veterinary hand-held glucometers. Frontiers in Veterinary Science, 12. https://doi.org/10.3389/fvets.2025.1694081

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free