Abstract
Background: In this study, the tear resistance of porcine lens capsules after continuous curvilinear capsulorhexis (CCC) and femtosecond (fs)-laser-assisted capsulotomy for cataract surgery (FLC) with different laser parameters is measured with a custom-made testing setup. Methods: Forty-five fresh porcine lenses were randomly chosen for CCC (n = 15) or FLC 1 (n = 15) and FLC 2 (n = 15). The FLC 1-group was treated with smaller spot distances than the FLC 2-group. The force necessary to break the opening of the anterior capsule and the maximum displacement were measured. Results: The mean tear resistance of the CCC-group (150 ± 70 mN) was higher than that of the FLC 1-group (60 ± 20 mN) and the FLC 2-group (30 ± 20 mN). Conclusion: It could be shown that CCC leads to a significantly higher tear resistance of the opening than FLC in porcine lenses. The femtosecond laser group demonstrated that smaller spot distances lead to a higher tear resistance.
Author supplied keywords
Cite
CITATION STYLE
Patzlaff-Günther, M., Fromm, M., Kern, T., Bartram, M., Schwäblein, A., Heinemann, D., … Ripken, T. (2022). Measurement of tear resistance after manual capsulorhexis and femtosecond laser–assisted capsulotomy of crystalline lenses. Lasers in Medical Science, 37(3), 1891–1897. https://doi.org/10.1007/s10103-021-03448-1