Abstract
The detection limits of matrix-assisted laser desorption/ionisation mass spectrometry (MALDI-MS) by semi-online coupling with chromatography were investigated using various mixtures of polyethylene oxides (PEOs) with different end groups. In contrast to the common 'dried-droplet' sample preparation technique, which results in an inhomogeneous sample-to-matrix ratio within the MALDI spot, the used coupling technique offers a very high reproducibility combined with surpassing sensitivity of a few femtograms over a broad range of sample-to-matrix ratios. These results are in correlation with the results of the recently established solvent-free MALDI-TOFMS method utilising the grinding approach and are also of assistance towards the more theoretical aspect of MALDI that suggests that there is no necessity for an analyte incorporation into a matrix crystal for excellent matrix-assistance. Copyright © 2005 John Wiley & Sons, Ltd.
Cite
CITATION STYLE
Falkenhagen, J., & Weidner, S. M. (2005). Detection limits of matrix-assisted laser desorption/ionisation mass spectrometry coupled to chromatography - A new application of solvent-free sample preparation. Rapid Communications in Mass Spectrometry, 19(24), 3724–3730. https://doi.org/10.1002/rcm.2256