Abstract
A simplified microbiological assay for vitamin B12 estimation, completed on microtitre plates using a colistin sulphate resistant strain of Lactobacillus leichmannii (NCIB 12519), and cryopreserved cultures is described. The new assay correlated well with a more conventional 'tube' assay and was not influenced by the presence of antibiotics in serum. Evaluation of assay performance showed excellent interassay and intra-assay precision with quantitative recovery of added cyanocobalamin over a wide range of additions (94.9%-102.1%). The advantages of short incubation time, easy reading, and minimal reagent costs make this assay an attractive option in the routine clinical laboratory and in research.
Cite
CITATION STYLE
Kelleher, B. P., & Broin, S. D. O. (1991). Microbiological assay for vitamin B12 performed in 96-well microtitre plates. Journal of Clinical Pathology, 44(7), 592–595. https://doi.org/10.1136/jcp.44.7.592