Microfluidic devices powered by integrated elasto-magnetic pumps

11Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

Abstract

We show how an asymmetric elasto-magnetic system provides a novel integrated pumping solution for lab-on-a-chip and point of care devices. This monolithic pumping solution, inspired by Purcell's 3-link swimmer, is integrated within a simple microfluidic device, bypassing the requirement of external connections. We experimentally prove that this system can provide tuneable fluid flow with a flow rate of up to 600 μL h-1. This fluid flow is achieved by actuating the pump using a weak, uniform, uniaxial, oscillating magnetic field, with field amplitudes in the range of 3-6 mT. Crucially, the fluid flow can be reversed by adjusting the driving frequency. We experimentally prove that this device can successfully operate on fluids with a range of viscosities, where pumping at higher viscosity correlates with a decreasing optimal driving frequency. The fluid flow produced by this device is understood here by examining the non-reciprocal motion of the elasto-magnetic component. This device has the capability to replace external pumping systems with a simple, integrated, lab-on-a-chip component.

Cite

CITATION STYLE

APA

Binsley, J. L., Martin, E. L., Myers, T. O., Pagliara, S., & Ogrin, F. Y. (2020). Microfluidic devices powered by integrated elasto-magnetic pumps. Lab on a Chip, 20(22), 4285–4295. https://doi.org/10.1039/d0lc00935k