Evaluation of Operator and Patient Doses after Irradiation with Handheld X-ray Devices

4Citations
Citations of this article
16Readers
Mendeley users who have this article in their library.

Abstract

This study aimed to evaluate radiation doses from handheld X-ray devices, specifically NOMAD Pro 2TM (DvcN), Rextar X (DvcRX), and Diox 602 (DvcD), targeting operator and patient’s critical organs and tissues. Calibrated TLD-100H dosimeters were placed on a mannequin and phantom head, focusing on the eyes, thyroid, gonads, hands, and salivary glands. Using a TLD reader, absorbed equivalent doses post irradiation were assessed. Conventional systems yielded higher radiation doses than phosphor plates and digital systems. Notably, implementing protective measures resulted in significant (p < 0.05) dose reductions to the operator. Peak measurements without protection included: gonad 24.4 (DvcN) μGy; thyroid 30.5 (DvcN) μGy; right eye 31.9 (DvcN) μGy; left eye 27.9 (DvcN) μGy; right hand 111.6 (DvcRX) μGy; and left hand 71.7 (DvcD) μGy. Radiation dose reductions ranged from 11.49% to 93.25%, depending on the region and device. It is imperative to adhere to radiological protection protocols, particularly when employing handheld X-ray devices; optimally, these should be used with digital systems.

Cite

CITATION STYLE

APA

Altındağ, A., Eren, H., Orhan, K., & Görgün, S. (2023). Evaluation of Operator and Patient Doses after Irradiation with Handheld X-ray Devices. Applied Sciences (Switzerland), 13(18). https://doi.org/10.3390/app131810414

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free