Analytical validation of two point-of-care assays for serum amyloid a measurements in cats

5Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

Abstract

Serum Amyloid A (SAA) is one of the most sensitive tests to detect inflammation in cats. In this study, two point-of-care assays for SAA measurements in cats (FUJI DRI-CHEM IMMUNO AU CARTRIDGE vf-SAA (method A), and CUBE-VET analyser (Method B), were analytically evaluated. Regarding the imprecision precision only the method A showed intra-assay and inter-assay CV < 10% at all concentrations. Both assays showed linearity with r close to 1 and the recovery were in the range of 81–112% for assay A and 85–125% for assay B and the limit of detection were 3.75 and 0.5 mg/dL for method A and B, respectively. A previously validated method for SAA quantification SAATIA; LZ-SAA (method C) was used as gold-standard to evaluate the accuracy of the assays. Significant correlations (p < 0.0001) were found between assays A and C (r = 0.94) and B and C (r = 0.91). In addition, an overlap performance test was made using serum samples from cats with non-inflammatory and cats with inflammatory. Both assays showed higher median SAA concentrations in cats with inflammatory diseases than in cats without inflammatory diseases (p < 0.0001). In conclusion, this manuscript provides data about the possible application of two point-of-care assays for the measurement of SAA concentration in cats.

Cite

CITATION STYLE

APA

Escribano, D., Bustillo, A. O., Marín, L. P., Rabasco, A. N., Herrera, P. R., Cerón, J. J., & Tvarijonaviciute, A. (2021). Analytical validation of two point-of-care assays for serum amyloid a measurements in cats. Animals, 11(9). https://doi.org/10.3390/ani11092518

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free