Abstract
Purpose: To compare the dose area products of pediatric VCUG in daily practice with the dramatically reduced official German diagnostic reference levels, which are based on selected data. Materials and Methods: 413 consecutive pediatric VCUG examinations were analyzed. Results: The mean dose area product of all examinations was 0.97 dGycm. This is below the lowest reference level that is valid for neonates. In 12 cases (5.6 %) the achieved dose area product was higher than the corresponding reference level. Conclusion: Using the available techniques for radiation protection, it is possible in the daily routine to meet the official diagnostic reference levels for children, which have been reduced by up to 80 %, even though these levels are based on a selected, possibly non-representative data set.
Author supplied keywords
Cite
CITATION STYLE
Born, M., Spiller, L., Bachour, H., Heydweiller, A., & Franke, I. (2013). Das MCU bei Kindern. Dosiswerte im klinischen Alltag im Hinblick auf die stark gesenkten diagnostischen Referenzwerte. RoFo Fortschritte Auf Dem Gebiet Der Rontgenstrahlen Und Der Bildgebenden Verfahren, 185(3), 262–267. https://doi.org/10.1055/s-0032-1325609