A trial of three non-invasive blood pressure monitors compared with invasive blood pressure assessment in atrial fibrillation and sinus rhythm

2Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Objective: To investigate the accuracy of three non-invasive blood pressure (BP) devices in atrial fibrillation (AF) compared with invasive arterial BP. Methods: One hundred patients aged 45-90 years, 63% male (50 in AF and 50 age matched controls in sinus rhythm [SR]) were identified with arterial lines measuring beat-to-beat BP fluctuation. Non-invasive BP measurements utilising the manual sphygmomanometer (MS), PulseCor R6.5 (PC) and automated sphygmomanometer (AS) were taken simultaneously with invasive BP in a randomised sequence. This was repeated three times in each patient. Results: In SR differences in systolic BP (SBP) for MS, AS and PC were −0.34 mm Hg (95% CI −2.31 to 1.63; P =.733), −3.80 mm Hg (95% CI −5.73 to −1.87; P =.0001) and −3.90 mm Hg (95% CI −5.90 to −1.90; P =.0001) and for diastolic BP (DBP) were 6.02 mm Hg (95% CI 4.39-7.64; P

Cite

CITATION STYLE

APA

Eysenck, W., Ammar, A., Kanthasamy, V., Freemantle, N., Veasey, R. A., Patel, N. R., … Sulke, N. (2019). A trial of three non-invasive blood pressure monitors compared with invasive blood pressure assessment in atrial fibrillation and sinus rhythm. International Journal of Clinical Practice, 73(12). https://doi.org/10.1111/ijcp.13410