Quality assurance tests for the Gamma Knife® Icon™ image guidance system

17Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Introduction: The Gamma Knife® Icon™ comes with an image guidance system for tracking patient motion and correcting for inter- and intrafractional shifts, mainly used with frameless thermoplastic immobilization. The system consists of a cone-beam CT (CBCT) and a couch-mounted infrared camera (IFMM). We report our quality assurance program for Icon's image guidance system. Methods: The manufacturer-provided tool is used for daily checks of CBCT positional precision. Catphan® phantom is used for monthly checks of CBCT image qualities (uniformity, contrast to noise ratio (CNR), and spatial resolution) for the two acquisition presets (low-dose and high-quality presets). On a semi-annual schedule, we use a frame tool to check the agreement of CBCT-based and Frame-based stereotactic space coordinates by comparing the locations of five attached ball bearings in CT-sim scans (Frame-based coordinates determination) and in Icon's CBCT scans. On an annual basis, the accuracy of IFMM, image registration, and delivery-after-shift are tested using a translational stage. A weighted CT dose index is measured annually with a pencil chamber in CTDI head phantom. Results: The CBCT precision check: 0.12 ± 0.04 mm (maximum deviations average). CBCT image quality: spatial resolution range: [6,7] lp/cm (low), and [7,8] lp/cm (high); uniformity: 12.82 ± 0.69% (low), and 13.01 ± 0.69% (high); CNR: 1.07 ± 0.08 (low), and 1.69 ± 0.10 (high). Agreement of CBCT-based with Frame-based stereotactic coordinates range: [0.33, 0.66] mm. Accuracy of IFMM: 0.00 ± 0.12 mm (average) with 0.27 mm (max.); image registration: 0.03 ± 0.06 mm (average) with 0.23 mm (max.); and delivery-after-shift: 0.24 ± 0.09 mm (average) with 0.42 mm (max.). CTDIw: 2.3 mGy (low), and 5.7 mGy (high). Conclusions: The manufacturer-required QA checks together with additional user-defined checks are an important combination for a robust quality assurance program ensuring the safe use of Gamma Knife® Icon™'s image guidance and motion management features.

Cite

CITATION STYLE

APA

AlDahlawi, I., Prasad, D., & Podgorsak, M. B. (2018). Quality assurance tests for the Gamma Knife® IconTM image guidance system. Journal of Applied Clinical Medical Physics, 19(5), 573–579. https://doi.org/10.1002/acm2.12417

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free