Accuracy of Breath-hold CT in Treatment Planning for Lung Stereotactic Ablative Radiotherapy

  • Mathieu D
  • Martel C
  • Campeau M
  • et al.
N/ACitations
Citations of this article
7Readers
Mendeley users who have this article in their library.

Abstract

PURPOSE The objectives of this study are (1) to measure concordance of tumor position on breath-hold (BH) computed tomography (CT) scans relative to the natural tumor path during free breathing (FB) and (2) to evaluate the benefits of the breathing monitoring device Abches (Apex Medical, Tokyo) for stereotactic ablative radiotherapy (SABR) treatment planning. METHODS In 53 lung cancer patients treated with CyberKnife™ robotic radiosurgery system, FB four-dimensional computerized tomography (4DCT) and end-expiration (EE) BH CT images were obtained. Extent of natural tumor motion was assessed with rigid registration derived from end-inspiration (EI) and EE phases of the 4DCT. Tumor displacement in BH scans relative to the natural tumor path was measured relative to the EE 4DCT phase. RESULTS Mean tumor motion (+/- 1 SD) during natural FB was 1 ± 1 mm, 2 ± 2 mm, and 6 ± 6 mm in medio-lateral, anterior-posterior, and cranio-caudal directions, respectively. Tumor position on BH CT scan was closer to EE than EI 4DCT phase for 35/53 patients (66%). Difference of BH tumor position vs. EE state was 4 ± 3 mm. Gross tumor displacements perpendicular to natural tumor path were as great as 11 mm (anterior-posterior) and were seen with or without the breathing monitoring device. CONCLUSION Tumor position during BH CT may not accurately correspond to positions observed on FB 4DCT. Hence, accurate and custom 4D analysis for each individual patient is recommended for treatment planning, especially those involving BH acquisitions.

Cite

CITATION STYLE

APA

Mathieu, D., Martel, C., Campeau, M.-P., Filion, É., Roberge, D., Bahig, H., … Carrier, J.-F. (2014). Accuracy of Breath-hold CT in Treatment Planning for Lung Stereotactic Ablative Radiotherapy. Cureus. https://doi.org/10.7759/cureus.236

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free