Abstract
The aim of this study was to verify the efficiency of the BAX® system for the detection of Salmonella spp. in raw chicken meat. The conventional culture method (IN 62, MAP) was used as a reference method. A total of 8,813 chicken carcass samples were analyzed. In the first part of the study, 1,200 samples were analyzed using the BAX® System and the conventional culture method. In the second part, 7,613 samples were analyzed by the BAX® system, and the conventional method was used only for samples that tested positive for Salmonella spp. by the BAX® system. The sensitivity, specificity, relative accuracy, positive predictive value, and negative predictive value obtained in the first part of this study were 100%, 92.3%, 96.4%, 53.3% and 100%, respectively. The BAX® system showed no false-negative results and reduced the time to obtain presumptive positive results. It is a suitable method for use in laboratories that perform a large number of food samples analyses daily. However, the conventional method is still required to confirm the presence of Salmonella spp. in samples that test positive using the BAX® system.
Author supplied keywords
Cite
CITATION STYLE
Frausto, H. S. E. G., Alves, J., & de Oliveira, T. C. R. M. (2013). Evaluation of the bax® system for the detection of salmonella spp. in naturally contaminated chicken meat. Food Science and Technology, 33(3), 475–478. https://doi.org/10.1590/S0101-20612013005000056