Abstract
Radiation doses to patients during cardiological procedures are of concern in the present day scenario. This study was intended to audit the radiation dose imparted to patients during the balloon mitral valvuloplasty (BMV) procedure. Thirty seven patients who underwent the BMV procedure performed using two dedicated cardiovascular machines were included in the study. The radiation doses imparted to patients were measured using a dose area product (DAP) meter. The mean DAP value for patients who underwent the BMV procedure from one machine was 19.16 Gy cm2 and from the other was 21.19 Gy cm2. Optimisation of exposure parameters and radiation doses was possible for one machine with the use of appropriate copper filters and optimised exposure parameters, and the mean DAP value after optimisation was 9.36 Gy cm 2. © 2006 IOP Publishing Ltd.
Cite
CITATION STYLE
Livingstone, R. S., Chandy, S., Timothy Peace, B. S., George, P., John, B., & Pati, P. (2006). Audit of radiation dose during balloon mitral valvuloplasty procedure. Journal of Radiological Protection, 26(4). https://doi.org/10.1088/0952-4746/26/4/004
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.