Abstract
Here, a novel biosensing platform for the detection of SARS-CoV-2 usable both at voltammetric and impedimetric mode is reported. The platform was constructed on a multi-walled carbon nanotubes (MWCNTs) screen-printed electrode (SPE) functionalized by methylene blue (MB), antibodies against SARS-CoV-2 spike protein (SP), a bioactive layer of chitosan (CS) and protein A (PrA). The voltammetric sensor showed superior performances both in phosphate buffer solution (PBS) and spiked-saliva samples, with LOD values of 5.0±0.1 and 30±2.1 ng/mL, compared to 20±1.8 and 50±2.5 ng/mL for the impedimetric sensor. Moreover, the voltammetric immunosensor was tested in real saliva, showing promising results.
Author supplied keywords
Cite
CITATION STYLE
Tortolini, C., Angeloni, A., & Antiochia, R. (2023). A Comparative Study of Voltammetric vs Impedimetric Immunosensor for Rapid SARS-CoV-2 Detection at the Point-of-care. Electroanalysis, 35(4). https://doi.org/10.1002/elan.202200349
Register to see more suggestions
Mendeley helps you to discover research relevant for your work.