An approach to the reduction of hydrocarbon contamination in the scanning electron microscope

58Citations
Citations of this article
81Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The deposition of electron beam-induced specimen contamination in both the transmission (TEM) and scanning electron microscopes (SEM) has remained a problem since the beginning of these forms of microscopy. Generally, sources of SEM contamination can be attributed to one or a combination of three major contributors: (1) the pumping system; (2) outgassing of other internal SEM component parts (i.e., specimen stage, stage lubricants, O-rings, etc.), or (3) the sample (including its preparation and handling). Generally, because of the nature of SEM, specimen contamination can be minimized but is difficult to avoid fully. This work outlines three approaches taken with instruments at NIST to reduce the deposition of contamination in high-resolution cold-field emission SEMs. With some modification these techniques could be applied to any SEM. These approaches have been in successful operation for several years, resulting in a reduction in electron beam-induced hydrocarbon contamination.

Cite

CITATION STYLE

APA

Postek, M. T. (1996). An approach to the reduction of hydrocarbon contamination in the scanning electron microscope. Scanning, 18(4), 269–274. https://doi.org/10.1002/sca.1996.4950180402

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley