In in this study ultrasonic method was applied for the measurement of blood pressure (BP). First, a novel method is proposed to measure mean arterial pressure (MAP), diastolic blood pressure (DBP) and systolic blood pressure (SBP) using ultrasonic sensors. The proposed algorithm is implemented by measuring the diameter of the artery and the speed of blood flow based on Doppler physical phenomena so that the BP can be calculated. The results of the proposed algorithm for MAP, SBP, and DBP hypertension analyses were evaluated with the results of Association for the Advancement of Medical Instrumentation (AAMI standard) for all three cases and their mean error rate for the worst case was -0.233mmHg and the standard deviation for 422 samples taken from individuals in the worst case was 4.53 mmHg that meets the standard requirements. Also, according to the British Hypertension Society (BHS) standard, the proposed algorithm for the estimation of BP for all three cases of MAP, DBP, and SBP has Grade A, indicating its higher accuracy in measuring and using the most effective variables in the diagnosis of hypertension in the human body. The proposed algorithm in BP estimation is non-invasive, cuff-less which needs no calibration, and is only based on using the ultrasonic sensor.
CITATION STYLE
Abolhasani, A., Mousazadeh, M., & Khoei, A. (2020). Real-time, cuff-less and non-invasive blood pressure monitoring. Informacije MIDEM, 50(2), 87–103. https://doi.org/10.33180/InfMIDEM2020.202