Abstract
In this paper the various approaches of EPID-based in vivo IMRT and VMAT dose verification, and their clinical implementation, are described. It will be shown that EPID-based in vivo dosimetry plays an important role in the total chain of verification procedures in a radiotherapy department. EPID-based dosimetry, in combination with in-room imaging, is a fast and accurate tool for 3D in vivo verification of VMAT delivery. EPID-based in vivo dosimetry provides clinically more useful information and is less time consuming than patient-specific pre-treatment dose verification. In addition to accurate 3D dose verification, in vivo EPID-based dosimetry will also detect major errors in the dose received by individual patients, and provides a safety net for advanced treatments such as IMRT and VMAT.
Cite
CITATION STYLE
Mijnheer, B., Olaciregui-Ruiz, I., Rozendaal, R., Sonke, J. J., Spreeuw, H., Tielenburg, R., … Mans, A. (2013). 3D EPID-based in vivo dosimetry for IMRT and VMAT. In Journal of Physics: Conference Series (Vol. 444). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/444/1/012011