Abstract
We performed calibrations of four different commercial picoammeters using the ultrastable low-noise current amplifier (ULCA) as a calibrator current source operated in the range between 1 fA and 1~\mu \text{A}. The results allow the comprehensive characterization of the devices under test regarding noise, settling, and burden voltage behavior as well as stability of the gain factor, and confirm the performance of the ULCA for use as small-current calibrator standard. Also, we present a further advanced setup for the calibration of transimpedance amplifiers. Accuracy limits for the best electrometer calibrations in the current range between 1 fA and 1~\mu \text{A} and possible implications on corresponding calibration and measurement capabilities are discussed.
Author supplied keywords
Cite
CITATION STYLE
Scherer, H., Drung, D., Krause, C., Gotz, M., & Becker, U. (2019). Electrometer Calibration with Sub-Part-Per-Million Uncertainty. IEEE Transactions on Instrumentation and Measurement, 68(6), 1887–1894. https://doi.org/10.1109/TIM.2019.2900129