Protein microdeposition using a conventional ink-jet printer

150Citations
Citations of this article
45Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Many recent bioanalytical systems based on immunologic and hybridization reactions in a mono- or bidimensional microarray format require technology that can produce arrays of spots containing biospecific molecules. Some microarray deposition instruments are commercially available, and other devices have been described in recent papers. We describe a system obtained by adapting a commercial ink-jet printer and used to produce mono- and bidimensional arrays of spots containing horseradish peroxidase on cellulose paper. In a few minutes, it was possible to obtain bidimensional arrays containing several thousands of spots with a diameter as low as 0.2 mm, with each of which requiring only a few nanoliters of the enzyme deposition solution. The quantity of enzyme in each spot was evaluated with a chemiluminescent reaction and a charge-coupled device-based, low-light imaging luminograph. The chemiluminescence measurements revealed that the reproducibility of the enzyme deposition was satisfactory for analytical purposes, with the variation coefficients being lower than 10% in almost all cases.

Cite

CITATION STYLE

APA

Roda, A., Guardigli, M., Russo, C., Pasini, P., & Baraldini, M. (2000). Protein microdeposition using a conventional ink-jet printer. BioTechniques, 28(3), 492–496. https://doi.org/10.2144/00283st07