Radiological properties of a wax-gypsum compensator material

4Citations
Citations of this article
9Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In this paper the radiological properties of a compensator material consisting of wax and gypsum is presented. Effective attenuation coefficients (EACs) have been determined from transmission measurements with an ion chamber in a Perspex phantom. Measurements were made at 80 and 100 cm source-to-skin distance (SSD) for beam energies of 6, 8, and 15 MV, for field sizes ranging from narrow beam geometries up to 40 × 40 cm2, and at measurement depths of maximum dose build-up, 5 and 10 cm. A parametrization equation could be constructed to predict the EAC values within 4% uncertainty as a function of field size and depth of measurement. The EAC dependence on off-axis position was also quantified at each beam energy and SSD. It was found that the compensator material reduced the required thickness for compensation by 26% at 8 MV when compared to pure paraffin wax for a 10 × 10 cm 2 field. Relative surface ionization (RSI) measurements have been made to quantify the effect of scattered electrons from the wax-gypsum compensator. Results indicated that for 80 cm SSD the RSI would exceed 50% for fields larger than 15 × 15 cm2. At 100 cm SSD the RSI values were-below 50% for all field sizes used. © 2005 American Association of Physicists in Medicine.

Cite

CITATION STYLE

APA

Du Plessis, F. C. P., & Willemse, C. A. (2005). Radiological properties of a wax-gypsum compensator material. Medical Physics, 32(5), 1246–1255. https://doi.org/10.1118/1.1896455

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free