Abstract
Present paper reports multianalyte electroanalysis of hypertensive having amlodipine and nebivolol molecules in a single dose. Amlodipine and Nebivolol exhibited oxidation peaks at 0.70 and 0.90 V respectively on the Graphite/Polyaniline-Cerium Oxide/Carbon paste electrode. The linearity range was 0.1 to 1.6 μ g ml −1 in Britton Robinson buffer (pH 8.0). The limit of detection was 20.0 ng ml −1 for Amlodipine and 30.0 ng ml −1 for Nebivolol and the limit of quantification was 80.0 ng ml −1 for Amlodipine and 100 ng ml −1 for Nebivolol respectively. These analytes were also determined in pharmaceutical formulation and human serum and good recovery was obtained for the developed method.
Cite
CITATION STYLE
Sharma, S., Jain, R., & Pandey, A. (2022). Advanced Sensing Performance towards Simultaneous Determination of Binary Mixture of Antihypertensives Using PANI–Cerium Oxide Nanoparticles as Modifier in Carbon Paste Incorporating Graphite and Silicon-Oil. Journal of The Electrochemical Society, 169(6), 066511. https://doi.org/10.1149/1945-7111/ac76e2