High clearance continuous renal replacement therapy with a modified dialysis machine

63Citations
Citations of this article
26Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Recent studies suggest that the dialysis dose significantly affects survival in acute renal failure (ARF) patients and that bicarbonate dialysate improves acid-base balance during continuous renal replacement therapy (CRRT). These data inspired us to use slow continuous dialysis (SCD) in the treatment of ARF. SCD is defined by the following parameters: (a) blood flow (Q(B)) = 100 to 200 ml/min, (b) dialysate flow (Q(D)) = 100 to 300 ml/min, (c) the use of a modified hemodialysis machine with controlled ultrafiltration and online production of bicarbonate-based dialysate, and (d) continuous or extended daily treatment for 8 to 24 hours SCD provides a urea clearance in the 70 to 80 ml/min range. Preliminary data from an ongoing clinical trial demonstrate the safety, efficiency, and simplicity of the treatment.

Cite

CITATION STYLE

APA

Schlaeper, C., Amerling, R., Manns, M., & Levin, N. W. (1999). High clearance continuous renal replacement therapy with a modified dialysis machine. Kidney International, Supplement, 56(72). https://doi.org/10.1046/j.1523-1755.56.s72.13.x

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free