Heart rate control using first- and second-order models was compared using a novel control design strategy which shapes the input sensitivity function. Ten participants performed two feedback control test series on a treadmill with square wave and constant references. Using a repeated measures, counterbalanced study design, each series compared controllers C1 and C2 based on first- and second-order models, respectively. In the first series, tracking accuracy root-mean-square tracking error (RMSE) was not significantly lower for C2: 2.59 bpm vs. 2.69 bpm (mean, C1 vs. C2), p = 0.79. But average control signal power was significantly higher for C2: (Formula presented.) vs. (Formula presented.), (Formula presented.). In the second series, RMSE was also not significantly lower for C2: 1.99 bpm vs. 1.94 bpm, p = 0.39; but average control signal power was again significantly higher for C2: (Formula presented.) vs. (Formula presented.), p = 0.045. The results provide no evidence that controllers based on second-order models lead to better tracking accuracy, despite the finding that they are significantly more dynamic. Further investigation using a substantially larger sample size is warranted.
CITATION STYLE
Wang, H., & Hunt, K. J. (2021). Heart rate control using first- and second-order models during treadmill exercise. Systems Science and Control Engineering, 9(1), 651–662. https://doi.org/10.1080/21642583.2021.1976304