Treatment of EXAFS data taken in the fluorescence mode in non-linear conditions

53Citations
Citations of this article
17Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The aim of this work is to investigate the possibility of extracting correct structural parameters from fluorescence EXAFS data taken at high count rates with an energy-resolving detector. This situation is often encountered on third-generation synchrotron radiation sources which provide a high flux on the sample. Errors caused by pulse pileup in the extraction of structural information have been quantified in a real experiment, and different approaches to the problem of data correction have been elaborated. The different approaches are discussed in a comparison of the ability of each kind of correction to recover the correct structural parameters. The result of our analysis is that it is possible to work in non-linear conditions and correct the data, if the response of the acquisition system is known. Reliable structural information can be obtained with data acquired up to a count rate equal to approximately 60% of the inverse of the dead time.

Cite

CITATION STYLE

APA

Ciatto, G., D’Acapito, F., Boscherini, F., & Mobilio, S. (2004). Treatment of EXAFS data taken in the fluorescence mode in non-linear conditions. Journal of Synchrotron Radiation, 11(3), 278–283. https://doi.org/10.1107/S0909049504002122