Electrically evoked whole-nerve action potentials: Parametric data from the cat

  • Brown C
  • Abbas P
72Citations
Citations of this article
39Readers
Mendeley users who have this article in their library.
Get full text

Abstract

In a companion paper [Brown et al., J. Acoust. Soc. Am. 88, 1385–1391 (1990)], a method for recording the electrically evoked whole-nerve action potential in human cochlear implant users was reported. The procedure for recording the response requires that two biphasic current pulses, a ‘‘masker’’ and a ‘‘probe,’’ be presented at a rate and level sufficient to drive the auditory nerve into a refractory state. The present study was designed to assess the sensitivity of that recording technique to variations in stimulation parameters. The experiments described in this paper demonstrate that: (1) the EAP as recorded in the cat is triphasic and is defined by two negative peaks occurring at latencies of approximately 0.26 and 0.82 ms; (2) EAP amplitude is independent of the level of the masker stimulus for current levels equal to or greater than the current level of the probe stimulus; and (3) the time course of recovery of the EAP from the refractory state is stable over a range of both probe and masker current levels.

Cite

CITATION STYLE

APA

Brown, C. J., & Abbas, P. J. (1990). Electrically evoked whole-nerve action potentials: Parametric data from the cat. The Journal of the Acoustical Society of America, 88(5), 2205–2210. https://doi.org/10.1121/1.400117