Abstract
When performing electrochemical impedance spectroscopy (EIS) measurements on a system, we must make sure it fulfills certain conditions. One of them is that it should be stationary that is to say, steady-state and time-invariant. Commonly studied systems are time-variant, for example a corroding electrode or a battery under operation. A corroding electrode sees its polarization resistance decrease with time. A passivating electrode sees its polarization resistance increase with time. These phenomena cause a deformation of the Nyquist impedance at low frequencies. This result was first simulated and validated by experimental measurements on a corroding steel sample undergoing uniform cor-rosion. The effect of performing impedance measurements on a discharging battery was also shown. Several methods are available to check and correct time-variance. The non-stationary distortion (NSD) indicator is used to separate valid and invalid data samples and the so called “4D impedance” method can easily produce instantaneous impedance data.
Author supplied keywords
Cite
CITATION STYLE
Murer, N., Diard, J. P., & Petrescu, B. (2020). The effects of time-variance on impedance measurements: Examples of a corroding electrode and a battery cell. Journal of Electrochemical Science and Engineering, 10(2), 127–140. https://doi.org/10.5599/jese.725