Calibration of area detectors from powder diffraction standards is widely used at synchrotron beamlines. From a single diffraction image, it is not possible to determine both the sample-to-detector distance and the wavelength, but, with images taken from multiple positions along the beam direction and where the relative displacement is known, the sample-to-detector distance and wavelength can both be determined with good precision. An example calibration using the GSAS-II software package is presented.
CITATION STYLE
Horn, C., Ginell, K. M., Von Dreele, R. B., Yakovenko, A. A., & Toby, B. H. (2019). Improved calibration of area detectors using multiple placements. Journal of Synchrotron Radiation, 26, 1924–1928. https://doi.org/10.1107/S1600577519013328