Temperature quantitation and mapping of frozen tissue

49Citations
Citations of this article
29Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A method was developed for quantitating the temperature within frozen tissue with the magnetic resonance (MR) parameter R2*. The pulse sequence uses half-pulse excitation and a short spiral readout to achieve echo times as short as 0.2 msec. Fiber-optic temperature sensors were inserted into bovine liver tissue. The tissue was frozen at one end while being held warm at the other end. Once steady state was reached, the parameter R2* was measured. A linear dependence of R2* on temperature was demonstrated. R2* is independent of freeze number and of the orientation of the temperature gradient with respect to the main magnetic field. Feasibility in a canine prostate during cryosurgery is demonstrated. © 2001 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Butts, K., Sinclair, J., Daniel, B. L., Wansapura, J., & Pauly, J. M. (2001). Temperature quantitation and mapping of frozen tissue. Journal of Magnetic Resonance Imaging, 13(1), 99–104. https://doi.org/10.1002/1522-2586(200101)13:1<99::AID-JMRI1015>3.0.CO;2-O