Detecting and removing data artifacts in Hadamard transform ion mobility-mass spectrometry measurements

48Citations
Citations of this article
38Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Applying Hadamard transform multiplexing to ion mobility separations (IMS) can significantly improve the signal-to-noise ratio and throughput for IMS coupled mass spectrometry (MS) measurements by increasing the ion utilization efficiency. However, it has been determined that fluctuations in ion intensity as well as spatial shifts in the multiplexed data lower the signal-to-noise ratios and appear as noise in downstream processing of the data. To address this problem, we have developed a novel algorithm that discovers and eliminates data artifacts. The algorithm employs an analytical approach to identify and remove artifacts from the data, decreasing the likelihood of false identifications in subsequent data processing. Following application of the algorithm, IMS-MS measurement sensitivity is greatly increased and artifacts that previously limited the utility of applying the Hadamard transform to IMS are avoided.

Cite

CITATION STYLE

APA

Prost, S. A., Crowell, K. L., Baker, E. S., Ibrahim, Y. M., Clowers, B. H., Monroe, M. E., … Payne, S. H. (2014). Detecting and removing data artifacts in Hadamard transform ion mobility-mass spectrometry measurements. Journal of the American Society for Mass Spectrometry, 25(12), 2020–2027. https://doi.org/10.1007/s13361-014-0895-y