Development and validation of RP-HPLC method for the estimation of alfuzosin hydrochloride in bulk drug and tablets

ISSN: 0019462X
1Citations
Citations of this article
20Readers
Mendeley users who have this article in their library.

Abstract

A reverse phase high performance liquid chromatography (RP-HPLC) method has been developed for the estimation of Alfuzosin hydrochloride in bulk drug and pharmaceutical dosage form. The quantification was carried out using a RP stainless steel column in isocratic mode, with mobile phase consisting of methanol, water, triethyl amine in the ratio of 60: 39.6: 0.4 (v/v) and added concentrated formic acid to make up the pH 4. The mobile phase was pumped at a rate of 0.9 mL/min and the detection was carried out at 245 nm and the linearity was found to be in the range of 10 to 1000 mcg/mL. The proposed method was found to be simple, precise, accurate, rapid and reproducible for the estimation of Alfuzosin hydrochloride in bulk drug and tablets.

Cite

CITATION STYLE

APA

Chandra, A., Channabasavaraj, K. P., Manohara, Y. N., & Ankita, S. M. (2007). Development and validation of RP-HPLC method for the estimation of alfuzosin hydrochloride in bulk drug and tablets. Indian Drugs, 44(7), 555–558.