Quantitative assessment of cardiac output and left ventricular function by noninvasive phase-contrast and cine MRI: Validation study with invasive pressure-volume loop analysis in a swine model

15Citations
Citations of this article
53Readers
Mendeley users who have this article in their library.
Get full text

Abstract

Purpose: To validate noninvasive cardiac output measurements of phase-contrast magnetic resonance imaging (PC-MRI) and cine MRI using an invasive pressure-volume (PV) loop technique on a swine model. Materials and Methods: We compared three methods for evaluating cardiac function at rest and under pharmaceutical low-dose inotropic infusion conditions: 1) phase-contrast MRI, 2) cine MRI, and 3) PV loop relationship. These measurements were made in 14 domestic pigs under rest conditions. Identical MRI acquisitions and PV loop analysis were performed on six pigs from the same group that received an infusion of dobutamine 2.5 μkg/min. Cardiac outputs from all measurements were analyzed and compared using linear regression and Bland-Altman analysis. Results: Noninvasive PC-MRI and cine MRI did not show any significant differences compared to an invasive PV loop technique for measurement of cardiac output under both rest (PC-MRI, cine MRI, and PV loop, 3.17 ± 0.45, 3.18 ± 0.61, 3.45 ± 0.41 L/min, respectively) and pharmaceutical low-dose inotropic infusion conditions (PC-MRI, cine MRI, and PV loop, 4.78 ± 0.53, 4.7 ± 0.6, 4.96 ± 0.48 L/min, respectively). Statistical analysis showed good agreement of cardiac output measurements at rest (R2 = 0.83) and under low-dose inotropic infusion conditions (R2 = 0.74) using PC-MRI and PV loop techniques. Cardiac output measurement using cine MRI and PV loop techniques also showed good agreement at rest (R2 = 0.85) and under low-dose inotropic infusion conditions (R2 = 0.76). Furthermore, cardiac outputs determined with the three modalities showed good agreement over a wide range of heart rates (90-180 bpm). Conclusion: MRI can provide a reliable, noninvasive measurement of cardiac output that can be carried out without the complications that are inherent with current invasive procedures. © 2011 Wiley-Liss, Inc.

Cite

CITATION STYLE

APA

Lin, H. Y., Freed, D., Lee, T. W. R., Arora, R. C., Ali, A., Almoustadi, W., … Tian, G. (2011). Quantitative assessment of cardiac output and left ventricular function by noninvasive phase-contrast and cine MRI: Validation study with invasive pressure-volume loop analysis in a swine model. Journal of Magnetic Resonance Imaging, 34(1), 203–210. https://doi.org/10.1002/jmri.22587

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free