Abstract
Purpose: To set up a robust and patient-friendly whole-heart protocol based on 32-receive-channel technology that will potentially allow a large part of the patient population to be addressed. Materials and Methods: Ten volunteers were examined on a clinical 1.5 T scanner equipped with a 32-channel data acquisition system using an experimental 32-element coil array. A magnetization-prepared, navigator-gated and -tracked 3D Cartesian balanced FFE sequence was used for whole-heart coronary MR angiography (MRA). With the use of sensitivity encoding (SENSE) and partial Fourier encoding for scan acceleration, nearly isotropic high-resolution data sets were acquired during free breathing in four minutes. Results: A high contrast and sufficient signal-to-noise ratio (SNR) were obtained, which allowed visualization of the major vessels up to the distal regions and detection of major branches. Phase encoding in the anterior-posterior (AP) direction was the most favorable SENSE configuration and allowed a reasonable scan time reduction with moderate SENSE factors. Conclusion: The employed 32-receive channel technology enabled a robust trade-off among SNR, spatial resolution, and scan time. In this study the most robust results were obtained using the smallest possible SENSE factors for a given voxel size and scan time. © 2006 Wiley-Liss, Inc.
Author supplied keywords
Cite
CITATION STYLE
Nehrke, K., Börnert, P., Mazurkewitz, P., Winkelmann, R., & Gräßlin, I. (2006). Free-breathing whole-heart coronary MR angiography on a clinical scanner in four minutes. Journal of Magnetic Resonance Imaging, 23(5), 752–756. https://doi.org/10.1002/jmri.20559