Abstract
Background: Transcatheter aortic valve-in-valve implantation (ViV TAVI) in degenerated Medtronic Freestyle aortic bioprosthesis (FSB) has been reported as being technically challenging. This study sought to evaluate procedural data and outcomes after ViV TAVI using a balloon-expandable Edwards valve in patients with failed FSB. Methods: Between August 2014 and December 2020, twenty-seven consecutive patients underwent ViV TAVI for symptomatic FSB failure at our institution using a Sapien XT (n=1) and Sapien 3 (n=26) valve, respectively. Endpoints were defined according to the Valve Academic Research Consortium-2 (VARC-2) criteria and were retrospectively analyzed. Results: Mean patient age was 75.7±8.2 years (female n=5, male n=22); Society of Thoracic Surgeons Predicted Risk of Mortality score was 7.3%±6.2%. ViV implantation with correct positioning of the Edwards Sapien valve within the FSB was successful in all cases. Intraprocedural transesophageal echocardiography revealed none/trace paravalvular regurgitation in twenty-five patients (92.6%), mild paravalvular regurgitation was present in two patients (7.4%). Neither of the patients had a mean gradient >20.0 mmHg excluding significant patient-prosthesis mismatch. Three early deaths (
Author supplied keywords
Cite
CITATION STYLE
Burgdorf, C., Momchilovska, A. V., & Remppis, B. A. (2021). Outcomes after transcatheter valve-in-valve implantation using a balloon-expandable Edwards Sapien valve in patients with degenerated Freestyle aortic bioprosthesis. Annals of Cardiothoracic Surgery, 10(5), 667–673. https://doi.org/10.21037/acs-2021-tviv-fs-43