Abstract
The collision/reaction interface with He introduced through the skimmer at 100 mL min-1 is employed to improve accuracy in 31P + determinations by ICP-QMS. The 31P16O + determination in cool plasma conditions result in limits of detection 55-fold lower. This strategy may be an interesting alternative for high ionization potential elements such as P. © 2012 Sociedade Brasileira de Química.
Author supplied keywords
Cite
CITATION STYLE
Donati, G. L., Amais, R. S., & Nóbrega, J. A. (2012). Strategies to improve accuracy and sensitivity in phosphorus determinations by inductively coupled plasma quadrupole mass spectrometry. Journal of the Brazilian Chemical Society, 23(4), 786–791. https://doi.org/10.1590/S0103-50532012000400026