Repeated hot water and steam disinfection of Pari LC Plus® nebulizers alter nebulizer output

7Citations
Citations of this article
21Readers
Mendeley users who have this article in their library.

This article is free to access.

Abstract

Currently, cystic fibrosis patients require daily nebulized treatments to achieve optimal lung health. Growth of pathogenic bacteria in patient nebulizers is well known, and disinfection guidelines have been established. In this short communication, we sought to discover what effect, if any, repeated nebulization/disinfection cycles had on nebulizer output. We nebulized saline repeatedly after exposure to boiling water, steam, and alcohol disinfection methods. While alcohol disinfection did not affect nebulizer output, boiling water and steam significantly decreased nebulizer output from baseline, 74.1 ± 5.9% (p = 0.022) and steam 63.6 ± 6.5% (p = 0.0048) after 60 cycles respectively. This decrease in nebulizer output could significantly increase the duration of nebulizer treatment time and negatively impact the burden of care on patients with cystic fibrosis.

Cite

CITATION STYLE

APA

Collins, M. S., O’Brien, M., Schramm, C. M., & Murray, T. S. (2019). Repeated hot water and steam disinfection of Pari LC Plus® nebulizers alter nebulizer output. Journal of Cystic Fibrosis, 18(2), 233–235. https://doi.org/10.1016/j.jcf.2018.08.005