Abstract
Cementless acetabular components require good initial fixation to allow bony in-growth. The initial press-fit stability, important for designs that do not rely on supplemental fixation, was examined for three designs with different porous coating systems: beads, fiber mesh, and plasma spray. The ability to withstand tangential loads was determined (maximum rim loads: 122-1730 N). The plasma-sprayed acetabular cups withstood the greatest tangential load within the specified range of motion of 150 μm (P<0.01). Differences in surface preparation of titanium acetabular cups may significantly affect the initial stability of the implants in rim loading.
Cite
CITATION STYLE
Markel, D. C., Hora, N., & Grimm, M. (2002). Press-fit stability of uncemented hemispheric acetabular components: A comparison of three porous coating systems. International Orthopaedics, 26(2), 72–75. https://doi.org/10.1007/s00264-001-0314-4