Fiber-Optic Ammonia Sensor for Measuring Synaptic Glutamate and Extracellular Ammonia

46Citations
Citations of this article
7Readers
Mendeley users who have this article in their library.
Get full text

Abstract

A fiber-optic ammonia gas sensor designed for neurochemIcal applications Is presented. Parameters evaluated in terms of effect on the steady-state and dynamic response of this sensor Include the indicator dye, concentrations of Indicator and total ammonia nitrogen In the Internal solution, volume of the Internal solution, structure of the gas-permeable membrane, and temperature. The final ammonia sensor responds over the concentration range from 7 to 3000 nM with a limit of detection of 7 nM and response times ranging from 2 to 5 mln. Glutamate oxidase Is Immobilized at the tip of this ammonia sensor to provide a glutamate biosensor with a detection limit of 0.1 μM when operated at pH 7.8. In addition, this ammonia sensor Is used to measure extracellular ammonia levels In perfused retinal and eye-cup tissue preparations. These measurements indicate a calcium-dependent, potassiumevoked release of ammonia during these depolarization conditions. © 1992, American Chemical Society. All rights reserved.

Cite

CITATION STYLE

APA

Kar, S., & Arnold, M. A. (1992). Fiber-Optic Ammonia Sensor for Measuring Synaptic Glutamate and Extracellular Ammonia. Analytical Chemistry, 64(20), 2438–2443. https://doi.org/10.1021/ac00044a024

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free