Abstract
A simple and flexible setup for conducting drug metabolism studies is described in this report. A heating block was designed for the Multimek liquid handler platform for incubation of multiple samples at 37°C in a 96-well format. This setup enables the rapid performance of drug metabolism experiments on a large number of samples. In this report, the authors present the validation of the system by 1) showing reproducible and consistent determination of the in vitro half-life of midazolam in every well across the entire plate and 2) determination of metabolic parameter values of midazolam, testosterone, diclofenac, warfarin, and dextromethorphan and inhibition parameter values of quinidine and ketoconazole, all comparable to literature values. In addition, the authors demonstrate the application of the setup to determining the metabolic stability of a set of proprietary compounds, the inhibition of activity of cytochrome P450 (CYP) enzymes, and the conduct of a single combination experiment that can simultaneously determine the metabolic stability and CYP inhibition activity. Overall, the system represents a simple, high-throughput and useful tool for drug metabolism screening in drug discovery. © 2003 The Society for Biomolecular Screening.
Author supplied keywords
Cite
CITATION STYLE
Saraswat, L. D., Caserta, K. A., Laws, K., Wei, D., Jones, S. S., & Adedoyin, A. (2003). A high-throughput method for enzyme kinetic studies. Journal of Biomolecular Screening, 8(5), 544–554. https://doi.org/10.1177/1087057103257290