Abstract
It is expected that both noise and activity distribution can have impact on the detectability of a myocardial defect in a cardiac PET study. In this work, we performed phantom studies to investigate the detectability of a defect in the myocardium for different noise levels and activity distributions. We evaluated the performance of three reconstruction schemes: Filtered Back-Projection (FBP), Ordinary Poisson Ordered Subset Expectation Maximization (OP-OSEM), and Point Spread Function corrected OSEM (PSF-OSEM). We used the Channelized Hotelling Observer (CHO) for the task of myocardial defect detection. We found that the detectability of a myocardial defect is almost entirely dependent on the noise level and the contrast between the defect and its surroundings. © 2014 Mananga et al.
Cite
CITATION STYLE
Mananga, E. S., El Fakhri, G., Schaefferkoetter, J., Bonab, A. A., & Ouyang, J. (2014). Myocardial defect detection using PET-CT: Phantom studies. PLoS ONE, 9(2). https://doi.org/10.1371/journal.pone.0088200
Register to see more suggestions