Abstract

In view of recent reports of contamination of the fibreoptic bronchoscope by tuberculous and non-tuberculous mycobacteria, we evaluated the disinfecting properties of aqueous glutaraldehyde (Cidex) and an iodophor (Prepodyne) against Mycobacterium tuberculosis and M gordonae. We found that a 15 minute disinfection procedure with either agent, coupled with initial vigorous mechanical cleaning of the bronchoscope and its accessories, is a quick and reliable method for preventing the contamination of the bronchoscope with mycobacteria.

Cite

CITATION STYLE

APA

Davis, D., Bonekat, H. W., Andrews, D., & Shigeoka, J. W. (1984). Disinfection of the flexible fibreoptic bronchoscope against Mycobacterium tuberculosis and M gordonae. Thorax. BMJ Publishing Group. https://doi.org/10.1136/thx.39.10.785