Abstract
An experimental investigation was undertaken to isolate and quantitatively determine the effect relative humidity has on the light-scattering ability of aerosols. Both the naturally-occurring ambient aerosol of State College, Pa., and several common test aerosols were used. A measured flow of aerosol was mixed with a measured flow of particle-free air to form a mixture of constant contaminant level; the humidity of this mixture was varied by controlling the moisture content of the clean diluent air. The total light scattered by a given aerosol sample, at various relative humidities, was measured with a Sinclair-Phoenix aerosol photometer (measures the total light scattered in the near forward direction). All measurements were carried out at atmospheric pressure, and after the particulates had an average of 1½ minutes to reach equilibrium with the water vapor. Natural and laboratory-generated aerosols were both tested in this manner. © 1969 Air & Waste Management Association.
Cite
CITATION STYLE
Lundgren, D. A., & Cooper, D. W. (1969). Effect of humidify on light-scattering methods of measuring particle concentration. Journal of the Air Pollution Control Association, 19(4), 243–247. https://doi.org/10.1080/00022470.1969.10466482