Development of a GEM based detector for the CBM Muon Chamber (MUCH)

13Citations
Citations of this article
5Readers
Mendeley users who have this article in their library.
Get full text

Abstract

The characteristics of triple GEM detectors have been studied systematically by using cosmic ray muons. The minimum ionizing particle (MIP) spectra have been taken for different GEM voltage setting. Efficiency of GEM detector has been measured for cosmic ray. At high rate operation of GEMs the value of the protection resistor influences the gain and the stability. This feature has been investigated varying both the rate and the value of the protection resistor. This measurement has been performed using both X-ray generator and Fe55 source. The ageing and long-term stability of GEM based detectors has been studied employing both X-ray generator and Fe 55 source. The ageing study of one GEM module is performed by using a 8 keV Cu X-ray generator to verify the stability and integrity of the GEM detectors over a longer period of time. The accumulated charge on the detector is calculated from the rate of the X-ray and the average gain of the detector. The details of the measurement and results has been described in this article.© 2013 IOP Publishing Ltd and Sissa Medialab srl.

Cite

CITATION STYLE

APA

Biswas, S., Schmidt, D. J., Abuhoza, A., Frankenfeld, U., Garabatos, C., Hehner, J., … Wiechula, J. (2013). Development of a GEM based detector for the CBM Muon Chamber (MUCH). Journal of Instrumentation, 8(12). https://doi.org/10.1088/1748-0221/8/12/C12002

Register to see more suggestions

Mendeley helps you to discover research relevant for your work.

Already have an account?

Save time finding and organizing research with Mendeley

Sign up for free