Temperature profile of ex-vivo organs during radio frequency thermal ablation by fiber Bragg gratings

  • Palumbo G
  • Iadicicco A
  • Tosi D
  • et al.
41Citations
Citations of this article
44Readers
Mendeley users who have this article in their library.

Abstract

© 2016 The Authors. We report on the integration of fiber optic sensors with commercial medical instrumentation for temperature monitoring during radio frequency ablation for tumor treatment. A suitable configuration with five fiber Bragg grating sensors bonded to a bipolar radio frequency (RF) probe has been developed to monitor the area under treatment. A series of experiments were conducted on ex-vivo animal kidney and liver and the results confirm that we were able to make a multipoint measurement and to develop a real-time temperature profile of the area, with a temperature resolution of 0.1°C and a spatial resolution of 5 mm during a series of different and consecutive RF discharges.

Cite

CITATION STYLE

APA

Palumbo, G., Iadicicco, A., Tosi, D., Verze, P., Carlomagno, N., Tammaro, V., … Campopiano, S. (2016). Temperature profile of ex-vivo organs during radio frequency thermal ablation by fiber Bragg gratings. Journal of Biomedical Optics, 21(11), 117003. https://doi.org/10.1117/1.jbo.21.11.117003