Laser CT scanning of LET-independent BANG3-Pro2® polymer gel dosimeters has recently shown potential in proton dosimetry. However, raw materials' impurities impart some variability. This study aimed to validate a new method of compensating for this variability, and to validate the suitability of the improved dosimeter for patient-specific QA in pencil beam scanning (PBS) and IMPT. Six modifications of the BANG3-Pro2® gel dosimeter were analysed for their sensitivity to proton dose and to LET. One formulation was selected for a clinical QA feasibility study, in which one composite IMPT plan, two single-field IMPT plans, and one SFUD plan were delivered to identical gel phantoms. New commercial VOLQATM software (beta version) was used for data analysis. Both validations were successful.
CITATION STYLE
Cardin, A., Ding, X., Kassaee, A., Lin, L., Maryanski, M. J., & Avery, S. (2015). Validation of high-resolution 3D patient QA for proton PBS and IMPT using laser CT of improved polymer gel dosimeters. In Journal of Physics: Conference Series (Vol. 573). Institute of Physics Publishing. https://doi.org/10.1088/1742-6596/573/1/012017