Abstract
A newly designed balloon coronary angioplasty catheter that allows passive antegrade blood flow during balloon inflation (autoperfusion catheter) was compared with a standard balloon coronary angioplasty catheter. In a randomized sequence, inflations were performed for 3 min in the left circumflex coronary artery of 12 dogs with the standard catheter followed by the autoperfusion catheter or vice versa. During inflation with the standard catheter, the ST segment of standard limb lead II increased from -0.02 ± 0.03 mV to 0.39 ± 0.08 mV (p < .01 compared with the standard catheter). Thus, unlike the standard catheter, the autoperfusion catheter allows for inflations up to 3 min in duration without producing deleterious changes in the ST segment or severe reductions in regional myocardial blood flow.
Cite
CITATION STYLE
Turi, Z. G., Campbell, C. A., Gottimukkala, M. V., & Kloner, R. A. (1987). Preservation of distal coronary perfusion during prolonged balloon inflation with an autoperfusion angioplasty catheter. Circulation, 75(6), 1273–1280. https://doi.org/10.1161/01.CIR.75.6.1273