The uncertainty in time of particle detection within a scintillator detector, characterised by the coinci- dence time resolution (CTR), is explored with respect to the interaction position within the scintillator crystal itself. Electronic collimation between two scintillator detectors is utilised to determine the CTR with depth of interaction (DOI) for different materials, geometries and wrappings. Significantly, no rela- tionship between the CTR and DOI is observed within experimental error. Confinement of the interaction position is seen to degrade the CTR in long scintillator crystals by 10%. © 2014 Brown et al.
CITATION STYLE
Brown, M. S., Gundacker, S., Taylor, A., Tummeltshammer, C., Auffray, E., Lecoq, P., & Papakonstantinou, I. (2014). Influence of depth of interaction upon the performance of scintillator detectors. PLoS ONE, 9(5). https://doi.org/10.1371/journal.pone.0098177