This study presents the development of stannous chloride (SnCl2)-modified glass substrates for biomolecule immobilization and their application in fabricating sensor chips for label-free interferometric biosensors. The glass modification process was optimized, identifying a 5% SnCl2 concentration, a 45 min reaction time, and a 150 °C drying temperature as conditions for efficient protein immobilization. Based on the SnCl2-modified glass substrates and label-free spectral-phase interferometry, a biosensor was developed for the detection of aflatoxin B1 (AFB1)—a highly toxic and carcinogenic contaminant in agricultural products. The biosensor realizes a competitive immunoassay of a remarkable detection limit as low as 26 pg/mL of AFB1, and a five-order dynamic range. The biosensor performance was validated using real corn flour samples contaminated with Aspergillus flavus. The proposed approach not only provides a powerful tool for AFB1 detection for food safety monitoring but also demonstrates the potential of SnCl2-modified substrates as a versatile platform for the development of next-generation biosensors.
Author supplied keywords
Cite
CITATION STYLE
Orlov, A. V., Zolotova, M. O., Novichikhin, D. O., Belyakov, N. A., Protasova, S. G., Nikitin, P. I., & Sinolits, A. V. (2024). Stannous Chloride-Modified Glass Substrates for Biomolecule Immobilization: Development of Label-Free Interferometric Sensor Chips for Highly Sensitive Detection of Aflatoxin B1 in Corn. Biosensors, 14(11). https://doi.org/10.3390/bios14110531